A licensee must retain a record of each spot check for gamma stereotactic radiosurgery units required under part 4731.4474 for three years. The record must include:
the manufacturer's name, model number, and serial number for the gamma stereotactic radiosurgery unit;
the model number, serial number, and identity of the instrument used to measure the output of the unit;
notations indicating the operability of radiation monitors, helmet microswitches, emergency timing circuits, emergency off buttons, electrical interlocks, source exposure indicator lights, viewing and intercom systems, timer termination, treatment table retraction mechanisms, and stereotactic frames and localizing devices (trunnions);