A licensee must perform acceptance testing on the treatment planning system of teletherapy and gamma stereotactic computer systems according to published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:
the accuracy of the software used to determine sealed source positions from radiographic images; and
the accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.