Before the first medical use of a brachytherapy source, a licensee must have:
determined the source output or activity using a dosimetry system that meets the requirements of part 4731.4468, subpart 1;
used published protocols currently accepted by nationally recognized bodies to meet the requirements of this item.
A licensee may use measurements provided by the source manufacturer or by a calibration laboratory accredited by the American Association of Physicists in Medicine that are made according to item A.
A licensee must mathematically correct the outputs or activities determined under item A for physical decay at intervals consistent with one percent physical decay.